Time-resolved fluorescence detection kit for monitoring 5-fluorouracil plasma-drug concentration in real time
A time-resolved fluorescence and detection kit technology, applied in the fields of medical biotechnology and molecular diagnosis, can solve the problems of low sensitivity, long detection time, complicated and cumbersome operations, etc., and achieves no radioactive pollution, wide detection range and good repeatability. Effect
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0048] A time-resolved fluorescence detection kit for real-time monitoring of blood concentration of 5-fluorouracil, including calibrator 5-FU with a concentration of 0, 10, 50, 100, 750 and 1500ng / mL, and the composition is 2mg / ml BSA, 0.1% NaN 3 , 50mM pH 7.8 Tris-HCl buffer calibrator dilution; europium-labeled 5-FUA-OVA antigen; anti-5-FU mouse polyclonal antibody; components are 50Mm pH 7.8 Tris-HCl, 1.5% PEG6000, 0.2% BSA, 0.1% Proclin300, 0.02% bovine IgG, 0.1% Tween 20 and 0.84% NaCl assay buffer; composition is 1.25mmol / L pH 7.8 Tris-HCl, 2.5% Tween 20 and 21% NaCl 25× washing solution; composition is 15umol / L β-naphthoyl trifluoroacetone, 50umol / L trioctylphosphine oxide, 0.1% TritonX-100, 0.6% acetic acid, potassium hydrogen phthalate to adjust pH3.0~3.2 enhancement solution; goat anti-mouse IgG-coated reaction plate; also includes 3 coverslips and 1 copy of instructions.
[0049] Wherein, the preparation steps of europium-labeled 5-FUA-OVA antigen are as follow...
Embodiment 2
[0077]The methodological verification of the test kit of the present invention, according to the conventional manufacturing and verification procedures in the art, the test kit prepared above is tested, and the results are as follows:
[0078] (1) Analytical sensitivity and linear range: measure 10 times with the zero reference standard as the sample, and calculate its fluorescence value and standard deviation. The concentration obtained by substituting the fluorescence value obtained by subtracting 2 times the standard deviation from the mean value of the fluorescence measurement at this point into the standard equation is its analytical sensitivity. The analytical sensitivity of this reagent is determined to be 2.1ng / ml. The antigen was diluted into different concentrations for determination, and the linear range of the measured standard curve was 2.1-1500ng / ml, the standard curve was LOG(Y)=5.024-0.194LOG(X), R=0.9986, see the standard curve figure 2 .
[0079] (2) Precis...
Embodiment 3
[0086] Test kit of the present invention and liquid chromatography-mass chromatography (LC-MS / MS) clinical blood sample measurement value comparison: use test kit of the present invention and liquid chromatography (LC-MS / MS) to simultaneously test 110 parts of serum Samples are tested. Measure the 5-FU concentration result of the blood sample by the method of the present invention as the abscissa, and use the 5-FU concentration result measured by the LC-MS / MS method as the ordinate to do regression analysis, and the correlation equation is: Y=0.981X-3.267, the correlation coefficient r=0.981 ( image 3 ). The results of statistical processing show that the method of the present invention has a significant correlation with the measured value of the clinical blood sample by the LC-MS / MS method.
PUM
Property | Measurement | Unit |
---|---|---|
Linear | aaaaa | aaaaa |
Sensitivity | aaaaa | aaaaa |
Abstract
Description
Claims
Application Information
- R&D Engineer
- R&D Manager
- IP Professional
- Industry Leading Data Capabilities
- Powerful AI technology
- Patent DNA Extraction
Browse by: Latest US Patents, China's latest patents, Technical Efficacy Thesaurus, Application Domain, Technology Topic, Popular Technical Reports.
© 2024 PatSnap. All rights reserved.Legal|Privacy policy|Modern Slavery Act Transparency Statement|Sitemap|About US| Contact US: help@patsnap.com