Detection kit for human serum amyloid A protein
A technology of amyloid and detection kits, applied in the field of medical immunology, can solve the problems of increasing the amount of antibodies and production costs, loss of antibody binding force, false positive and false negative test results, etc., to improve accuracy and credibility , reduce the amount of antibody used, and reduce production costs
- Summary
- Abstract
- Description
- Claims
- Application Information
AI Technical Summary
Problems solved by technology
Method used
Image
Examples
Embodiment 1
[0026] Embodiment 1 Preparation of Human Serum Amyloid A Detection Kit
[0027] The kit of the present invention relates to the main raw materials of reagents as follows:
[0028] 1. Serum amyloid A antibody: the titer measured by indirect ELISA method is 1:100000.
[0029] 2. Latex: The present invention only exemplarily uses polystyrene latex particles with a diameter of 100-200 nm and carboxyl groups to carry out experiments.
[0030] The preparation of the main reagent of this embodiment is as follows:
[0031] Reagent R1: ammonium chloride buffer containing 1.5% PEG6000 (polyethylene glycol 6000), 0.8% NaCl. The reagent is a colorless transparent solution.
[0032] Reagent R2: latex particles were sensitized with anti-serum amyloid A antibody. The reagent is a milky white solution.
[0033] Specific steps are as follows:
[0034] 1. Take 1ml (100mg / ml) latex, wash 3 times with 0.1M (mol / L), pH5.0 MES solution (2-morpholineethanesulfonic acid buffer), and ultrasonica...
Embodiment 2
[0043] Embodiment 2 Determination of human serum amyloid A
[0044] Detection tool: Hitachi 7080 automatic analyzer.
[0045] Analysis method: two-point endpoint method; main wavelength: 546nm, secondary wavelength: 700nm: sample volume: 2.0ul; R1: 240ul; R2: 60ul; calibration method: Logit-log5p; reaction direction: rising; measurement temperature: 37°C; Add R2 when the sample is mixed with R1 for 5 minutes, read the absorbance A1 at the first point immediately, and read the absorbance A2 after 5 minutes. The reaction absorbance was calculated as the calibrated difference between A2 and A1.
[0046] Calculation method: multi-point calibration, the dose / response curve is made according to the absorbance and reference serum value, and the sample content can be calculated on the dose / response curve according to its absorbance value.
Embodiment 3
[0047] Example 3 Performance Evaluation of Human Serum Amyloid A Detection Kit
[0048] 1. Analytical Sensitivity Assessment
[0049] Use a 200mg / L sample to measure the kit, repeat the measurement 3 times, and convert the change in absorbance at the concentration of 198.9mg / L specified in the kit. The results show that the absorbance at the concentration of 198.9mg / L is 0.0730 (see Table 1).
[0050] Table 1
[0051] Measurement times Determination of the mean absorbance Absorbance at 198.9mg / L concentration 3 0.0734 0.0730
[0052] 2. High value linear evaluation
[0053] Mix the samples with SAA concentrations of 0mg / L and 400mg / L in different proportions, and repeat the measurement 3 times for each concentration (see Table 2)
[0054] The highest detection range of the detection kit of the present invention can reach 400mg / L, and the judgment basis is r2≥0.990.
[0055] Table 2
[0056] sample ratio Theoretical valuemg / L 3 times de...
PUM
Property | Measurement | Unit |
---|---|---|
The average particle size | aaaaa | aaaaa |
The average particle size | aaaaa | aaaaa |
Diameter | aaaaa | aaaaa |
Abstract
Description
Claims
Application Information
- R&D Engineer
- R&D Manager
- IP Professional
- Industry Leading Data Capabilities
- Powerful AI technology
- Patent DNA Extraction
Browse by: Latest US Patents, China's latest patents, Technical Efficacy Thesaurus, Application Domain, Technology Topic, Popular Technical Reports.
© 2024 PatSnap. All rights reserved.Legal|Privacy policy|Modern Slavery Act Transparency Statement|Sitemap|About US| Contact US: help@patsnap.com